본문 바로가기 주메뉴 바로가기

인류의 건강과 생명공학 발전을 위해 봉사하는 기업 드림셀

We Serving the Health and Biotechnology of Humanity

제품정보

> 제품정보 > 시약 > 해외 공식 대리점 > [CSB-E04605h] Human Interleukin 16,IL-16 ELISA KIT
제품명

[CSB-E04605h] Human Interleukin 16,IL-16 ELISA KIT

제품코드 : CSB-E04605h

제품정보 : Human Interleukin 16,IL-16

  • Sample Types: serum, plasma, cell culture supernates, tissue homogenates
  • Detect Range: 0.156 ng/mL-10 ng/mL
  • Sensitivity: 0.039 ng/mL

제품문의

  • 전화번호
    2292-8870

제품상세

Product Details

Description

This Human IL16 ELISA Kit was designed for the quantitative measurement of Human IL16 protein in serum, plasma, cell culture supernates, tissue homogenates. It is a Sandwich ELISA kit, its detection range is 0.156 ng/mL-10 ng/mL and the sensitivity is 0.039 ng/mL.

Target Nameinterleukin 16 (lymphocyte chemoattractant factor)
Alternative NamesHGNC:5980 ELISA Kit; HsT19289 ELISA Kit; IL-16 ELISA Kit; IL16 ELISA Kit; IL16_HUMAN ELISA Kit; Interleukin 16 precursor ELISA Kit; Interleukin-16 ELISA Kit; LCF ELISA Kit; Lymphocyte chemoattractant factor ELISA Kit; Neuronal interleukin 16 ELISA Kit; NIL16 ELISA Kit; prIL 16 ELISA Kit; PrIL16 ELISA Kit; Prointerleukin 16 ELISA Kit
AbbreviationIL16
Uniprot No.Q14005
SpeciesHomo sapiens (Human)
Sample Typesserum, plasma, cell culture supernates, tissue homogenates
Detection Range0.156 ng/mL-10 ng/mL
Sensitivity0.039 ng/mL
Assay Time1-5h
Sample Volume50-100ul
Detection Wavelength450 nm
Research AreaImmunology
Assay Principlequantitative
MeasurementSandwich
Precision
Intra-assay Precision (Precision within an assay): CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess.
Inter-assay Precision (Precision between assays): CV%<10%
Three samples of known concentration were tested in twenty assays to assess.
Linearity
To assess the linearity of the assay, samples were spiked with high concentrations of human IL-16 in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay.
 SampleSerum(n=4)
1:1Average %88
Range %83-97
1:2Average %97
Range %91-104
1:4Average %84
Range %81-88
1:8Average %89
Range %81-96
Recovery
The recovery of human IL-16 spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section.
Sample TypeAverage % RecoveryRange
Serum (n=5)10194-101
EDTA plasma (n=4)9490-98
Typical Data
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed.
ng/mlOD1OD2AverageCorrected
101.7741.8481.8111.644
51.0031.0141.0090.842
2.50.6660.6870.6770.510
1.250.4010.4120.4070.240
0.6250.2890.2870.2880.121
0.3120.2450.2530.2490.082
0.1560.2110.2240.2180.051
00.1630.1700.167 
ELISA Data AnalysisWatch ELISA data processing video & download Curve Expert if needed
Troubleshooting
and FAQs
ELISA kit FAQs
StorageStore at 2-8°C. Please refer to protocol.
Lead Time3-5 working days