제품명

[SEA022Mu] ELISA Kit for Endothelial protein C receptor (EPCR)

제품코드 : SEA022Mu

제품정보 : Signal transduction;CD & Adhesion molecule;Hematology;

  • Organism Species: Mus musculus (Mouse)
  • Sample Type: serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
  • Test Method: Double-antibody Sandwich
  • Assay Length: 3h
  • Detection Range: 78-5,000pg/mL
  • Sensitivity: The minimum detectable dose of this kit is typically less than 32pg/mL

제품문의

  • 전화번호
    2292-8870

제품상세

Specificity

This assay has high sensitivity and excellent specificity for detection of Endothelial protein C receptor (EPCR).
No significant cross-reactivity or interference between Endothelial protein C receptor (EPCR) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Endothelial protein C receptor (EPCR) and the recovery rates were calculated by comparing the measured value to the expected amount of Endothelial protein C receptor (EPCR) in samples.

MatrixRecovery range (%)Average(%)
serum(n=5)96-10499
EDTA plasma(n=5)95-105101
heparin plasma(n=5)79-10497

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Endothelial protein C receptor (EPCR) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Endothelial protein C receptor (EPCR) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Endothelial protein C receptor (EPCR) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample1:21:41:81:16
serum(n=5)96-105%90-104%82-96%95-103%
EDTA plasma(n=5)91-101%93-105%83-92%99-105%
heparin plasma(n=5)88-102%92-101%92-99%89-96%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

ReagentsQuantityReagentsQuantity
Pre-coated, ready to use 96-well strip plate1Plate sealer for 96 wells4
Standard2Standard Diluent1×20mL
Detection Reagent A1×120µLAssay Diluent A1×12mL
Detection Reagent B1×120µLAssay Diluent B1×12mL
TMB Substrate1×9mLStop Solution1×6mL
Wash Buffer (30 × concentrate)1×20mLInstruction manual1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.